Agilent Technologies U1253B User's And Service Manual page 174

Agilent u1253b true rms oled multimeter
Hide thumbs Also See for U1253B:
Table of Contents

Advertisement

6
Performance Tests and Calibration
Table 6-2 Performance verification tests
Step
Test function
11
Press
to select frequency counter
12
Press
to select divide-by-100 frequency
counter mode
13
Turn the rotary switch to the
14
Press
to select temperature measurement
[8]
15
Turn the rotary switch to the
16
Press
to select ACμA measurement
17
Turn the rotary switch to the
position
148
Range
[6]
999.99 kHz
99.999 MHz
[7]
position
10.000 nF
100.00 nF
1000.0 nF
10.000 μF
100.00 μF
1000.0 μF
10.000 mF
100.00 mF
–40 °C to
1372 °C
500 μA
position
5000 μA
500 μA
[1]
5000 μA
50 mA
440 mA
Reference
signals/values
Error limits
33250A output
200 mVrms,
± 52 Hz
100 kHz
600 mVrms, 10
± 5.2 kHz
MHz
5520A output
10.000 nF
± 108 pF
100.00 nF
± 1.05 nF
1000.0 nF
± 10.5 nF
10.000 μF
± 105 nF
100.00 μF
± 1.05 μF
1000.0 μF
± 10.5 μF
± 105 μF
10.000 mF
100.00 mF
± 3.1 mF
0 °C
± 1 °C
100 °C
± 2 °C
500 μA
± 0.3 μA
5000 μA
± 3 μA
500 μA, 1 kHz
± 3.7 μA
500 μA, 20 kHz
± 3.95 μA
5000 μA, 1 kHz
± 37 μA
5000 μA, 20 kHz
± 39.5 μA
± 80 μA
50 mA
400 mA
± 0.65 mA
U1253B User's and Service Guide