Performance Tests and Calibration
6
Charge terminals alert test
This test determines whether the charge terminal alert is
functioning properly.
Set the rotary switch to any position other than
,
,
or
.
Provide a voltage level more than 5 V to the
terminal.
An error message Error ON mA INPUT (as shown in
Figure 6-
3)
will be displayed on the secondary display, and a continuous
beep will persist until the positive lead is removed from the
terminal.
Figure 6-3 Charge terminal error message
Before conducting this test, make sure the beep function is not disabled in
N O T E
Setup.
U1253B User's and Service Guide
141