Figure 5-2 Self-Testing Time Display; Table 5-1 Battery Voltage And Corresponding Percentage Of Charg - Agilent Technologies U1253B User's And Service Manual

Agilent u1253b true rms oled multimeter
Hide thumbs Also See for U1253B:
Table of Contents

Advertisement

5
Maintenance
N O T E
Table 5-1 Battery voltage and corresponding percentage of charges in standby and charging modes
Condition
Trickle
Being charged up
126
4 Insert the red (+) and black (–) banana plugs (4 mm
plugs) of the DC adapter to the
terminals respectively. Ensure that the polarity of the
connection is correct.
The DC adapter can be replaced with a DC power supply set at DC 24 V
with an overcurrent limit of 0.5 A.
5 The display will show a countdown timer of 10 seconds
for the self- test to start. The multimeter will output short
single- tone sounds to remind you to charge the battery.
Press
to start charging the battery, or the
multimeter will automatically start charging after
10 seconds. It is recommended not to charge the battery
if the battery capacity is over 90%.

Figure 5-2 Self-testing time display

Battery voltage
6.0 V to 8.2 V
7.2 V to 10.0 V
and COM
Proportional percentage
0% to 100%
0% to 100%
U1253B User's and Service Guide