Performance Tests and Calibration
7

Charge Terminal Alert Test

This test determines if the charge terminal alert is operating
properly.
The meter sounds an alert when the
terminal
detects a voltage level of more than 5V but rotary switch is
not set to
position. The meter sounds an alert beep
and primary display indicates and flashes "Ch.Err" until the
lead is removed from
terminal.
Figure 56
Charge terminal alert
Before conducting this test, make sure the beep function is not disabled in
N O TE
setup.