Performance Tests and Calibration
6

Charge Terminal Alert Test

This test determines if the charge terminal alert is operating
properly.
The meter sounds an alert when the
terminal
detects a voltage level of more than 5 V and the rotary
switch is not set to the
position. The meter sounds
an alert beep and the primary display flashes the "Ch.Err"
until the lead is removed from the
terminal.
Figure 6-3 Charge terminal alert
Before conducting this test, make sure the beep function is not disabled at
N O TE
setup.
Agilent U1251B/U1252B User's and Service Guide
113