Testing Procedure

Step 1
Firmly insert the blood glucose test strip into the blood glucose test
meter port with the "3.3" facing towards you.
The blood glucose test meter automatically turns on displaying the
code, ambient temperature and the date and time.
The EVOLUTION
TM
will automatically recognize the blood glucose
test strip code number and adjust the blood glucose test meter
accordingly.
•A blinking blood glucose test strip will appear at the top of the screen indi-
cating the blood glucose test meter is ready for testing.
Note : Inserting the test strip in the wrong direction will turn the meter on, but you will
be unable to test.
CAUTION
1. If the code is not displayed after for a while, pull the blood glucose test strip out
of the port, and re-start the procedure from the beginning.
2. If the code on the LCD display window and on the test vial does not match, try
another new blood glucose test strip. If the mismatch persists, please contact
your local representative for help.
3. If the meter does not power on, pull the blood glucose test strip out of the port,
and re-insert the blood glucose test strip.
4. Avoid testing under direct sunlight, for a more accurate test result.
5. If you apply your blood sample too early, ERROR5 message will appear on
LCD ( refer to p35).
• For a more accurate test result avoid testing under direct sunlight.
22
www.infopia21.com
Testing Procedure
Correct
Completely Filled
Correct
Incorrect
Incorrect
Poorly filled
Sample size
Step 2
•Place the top edge of the blood glucose test strip to the area you
prepared for testing.
•The blood will automatically be "drawn" into the blood glucose test
strip channel.
•When the chamber has enough blood, your blood glucose test meter
will automatically count down.
CAUTION
1. If the countdown does not start, do not add more blood to the blood glucose test
strip! Discard the blood glucose test strip an restart testing.
2. If you do not conduct the test within 3 minutes, the blood glucose test meter will
automatically power off to save battery life. In this case the test procedure should
start again from the beginning.
Applying Blood
Evolution
blood glucose
TM
test meter only requires 0.3
sample volume.
23