Tm Blood Glucose Test Meter; Test Principle; Important Information - Infopia Evolution Operation Manual

Table of Contents

Advertisement

Dear EVOLUTION
TM
Blood Glucose Monitoring
System Owner
Thank you for choosing the EVOLUTION
System. This booklet has important information you must know about the
EVOLUTION
TM
Blood Glucose Monitoring System. Please read it carefully.
We understand that self-testing of blood glucose level provides a way to
control your diabetes and may give you peace of mind by testing regularly.
As a result EVOLUTION
TM
was developed to provide you with a fast and
accurate reading with a convenient and simple process. Our goal at
TM
EVOLUTION
is to provide the best quality healthcare products coupled
with superior customer service. Always consult with your healthcare profes-
sional before making any changes to your diabetes management. The
EVOLUTION
TM
Blood Glucose Monitoring System is for in vitro diagnostic
use only.

Important information

The EVOLUTION
TM
Blood Glucose Monitoring System is intended for use
outside the body (in vitro diagnostic use only). It should be used only for
testing blood glucose and only with fresh capillary whole blood samples. It
should not be used for the diagnosis of diabetes.
Consult your physician or diabetic healthcare professional about daily man-
agement of your diabetes and proper use of the glucometer.
Please pay close attention when handling blood. Improper procedures may
cause serious hazards to your health.
The EVOLUTION
TM
Blood Glucose Monitoring System contains small
parts. Please keep your monitoring system out of reach of children.
4
www.infopia21.com
TM
Blood Glucose Monitoring

Test Principle

The enzyme glucose oxidase on the blood glucose test strip reads specifi-
cally with the blood glucse. The current generated is converted and dis-
played as blood glucose value.
The
Blood Glucose Monitoring System is plasma-calibrated
EVOLUTION
TM
to allow easy comparison of results with laboratory methods.
Blood glucose meter which is calibrated against a whole blood method
may have different results in comparison to
The laboratory system used for calibration of the
Glucose Monitoring System is YSI 2300 STAT plus which is equipped with
a glucuse oxidase system.
Note
< Maintaining the EVOLUTION
1. Do not assemble or disassemble the blood glucose test meter.
2. It will have to keep in the place where the water does not reach.
3. Avoid getting dirt and moisture.
4. It does not have to keep in storage place of the chemical chemicals or
the place where gas etc occurs.
5. Keep out of direct sunlight.
CAUTION
1. If you do not check the code number, pull the blood glucose test
strip out, conduct procedures from beginning.
2. If the code of the blood glucose test meter and of the blood glucose
test strip vial does not match, you must set the code number.
3. If the code on the LCD window and on the blood glucose test vial
does not match, follow the coding procedure for accurate results.
.
EVOLUTION
TM
EVOLUTION
TM
blood glucose test meter >
Blood
TM
5