Testing Procedure
Correct : Completely filled
completely
filled
correct
Incorrect : Poorly filled
poorly filled
incorrect
Step 2
Place your fingertip to the top edge of finetest Auto-coding
blood on your finger will automatically draw into the applying channel of the
test strip. Allow your finger to remain until you hear a "beep" sound from the
Meter. A sample volume of 1.5μl is required.
finetest Auto-coding
TM
Caution
1. If the countdown does not start, do not add more blood to the test strip! Discard
the test strip an restart testing.
2. If you do not conduct the test within 3 minutes, the meter will automatically
power off to save battery life. In this case the test procedure should start again
from the beginning.
2 2
Step 3
After the beeping sound, the test will begin
automatically and your results will appear in 9
seconds. It should begin counting down from 9 to 1
second on the LCD display window. The LCD
window will display the result of your blood glucose
level, temperature and time.
Record the result value in your logbook. When test
strip is removed, the meter turns off automatically.
Test Strip. The
TM
Caution
1. If the test result is out of the test range, the Hi/Lo message will be shown on the
LCD. (refer to p34)
2. Safely discard used test strip and lancet, to a proper place, according to the
local regulation.