Testing Procedure - Infopia Finetest Auto-coding Premium Manual

Table of Contents

Advertisement

Available languages

Available languages

Step 1
Firmly insert the test strip into Meter test port while the meter is switched off. Insert down firmly
in direction of arrow on test strip (arrow up). Please do not insert the test strip upside down.
When you insert the test strip into the meter, the power automatically turns on with the code and
temperature. After 3 seconds, the code number and testing temperature will disappear.
When the strip sign blinks, apply your blood to the test port of strip.
Note : Inserting the test strip in the wrong direction will turn the meter on, but you will be unable
to test.
Caution
1. If the code is not displayed after 3 seconds, pull the test strip out of the port, and re-start
the procedure from the beginning.
2. If the code on the LCD display window and on the test vial does not match, try another new
strip. If the mismatch persists, please contact your local representative for help (contact
details: see last page).
3. If the meter does not power on, pull the test strip out of the port, and re-insert the test strip.
4. Avoid testing under direct sunlight, for a more accurate test result.
5. If you apply your blood sample too early, 'Er5' message will appear on LCD (refer to p. 71).

Testing Procedure

59