Start an acquisition

Once an experiment is committed, press Start on the main UI, enter the data file name and then
press OK to initiate the acquisition.
Note: The current experiment settings are saved in a temporary file, which is overwritten every time
the settings are modified. When SuperFACT is restarted, the last executed experiment is
automatically loaded into memory and it can be executed by pressing the Start command on the
main UI.
Important: If you modify and experiment and (without saving it) and then press Measure or Commit
followed by Start, the changes will take immediate effect, but the original experiment setting will
not be modified. It follows, that the changes will be lost when the original experiment is reloaded
into memory.

Check current experiment and modify default Advanced Settings

The current experiment can be inspected by pressing the See Experiment button. This control is
available even if an acquisition is underway.
The advanced settings can be accessed/modified/saved by pressing the Adv. Settings button.
On the Advanced SMU and Acquisition Settings UI, use the Default? button to save the current
settings as default. The Default setting will be loaded by pressing the Load Default button.
Note: Each experiment is associated with its own set of Advanced and SMU Settings, and these are
automatically loaded into memory with the experiment.