Connection Of Water Cooling Hoses - Hamamatsu Photonics C11440-22CU Instruction Manual

Table of Contents

Advertisement

C11440-22CU/C11440-22CU01 Instruction manual_Ver.1.4
8-2

CONNECTION OF WATER COOLING HOSES

(1) Place the camera on the stable table.
(2) Connect water cooling hose into the WATER connector on the camera.
• Insert the hose fully into the WATER connector on the camera. (as shown in Figure
8-3)
• Confirm the hose stops at it.
(3) Set the camera onto a microscope (If the camera is used on the microscope).
If it is easy to connect the hose onto the camera after the camera is set onto the
microscope then it is OK to connect the hose after the camera is set on the microscope.
(4) Connect the hose onto the circulating water cooler.
Follow the instruction on the circulating water cooler when you connect the hose onto
the circulating water cooler.
(5) Turn on the circulating water cooler and confirm the cooling water is flowing normally.
20
Water cooling hose
Figure 8-3
Stop the circulating water cooler when the water flow is abnormal
or water drop or splash is found.
WATER connector