Frame Rate Calculation; Read-Out Time Of The Horizontal Line - Hamamatsu Photonics C11440-22CU Instruction Manual

Table of Contents

Advertisement

C11440-22CU/C11440-22CU01 Instruction manual_Ver.1.4

10-3-2 FRAME RATE CALCULATION

[Calculation formula]
1/(Exp1+(Vn+10)×1H)
Vn
Exp1 = 9.7 µs to 10 s
1H
[Value of frame rate]
Horizontal width × Vertical width
Size and position for sub-array read-out can be configured according to the table below.

10-3-3 READ-OUT TIME OF THE HORIZONTAL LINE

Read-out time and exposure time can be varied with light sheet mode.
The range of exposure time with light sheet mode is as shown below.
The maximum exposure time can be decided according to the readout time of a frame, which is
longer than 10 s or not.
46
= Number of vertical line (the number of lines from the middle of the sensor.)
6
=
× 10
)
HLN / (266
(HLN = 2592 to 26600000)
The Exp1 value must be input to the calculation formula below in units of seconds.
2048
2048
1024
512
256
128
64
8
4
Settings
Size
Vertical 4 lines steps
Position
Readout time = Vn × 1H
Vn = Number of vertical line
1H = HLN / (266 × 10
(HLN = 2592 to 26 600 000)
1H to 1H × Vn (1H × Vn < 10 s)
1H to 10 s
Frame rate (frame/s)
49
99
196
385
743
1386
5701
7330
6
): 9.7436 µs to 100 ms
(1H × Vn ≧10 s)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

C11440-22cu01

Table of Contents