HELPFUL HINT:
• Glucose level alerts and alarms will not function after the sensor session
has ended.
8.2
REMOVING A SENSOR
Sensors may fracture on rare occasions. If a sensor breaks and no
portion of it is visible above the skin, do not attempt to remove it.
Seek professional medical help if you have symptoms of infection
or inflammation—redness, swelling or pain—at the insertion site.
If you experience a broken sensor, please report this to your local
distributor.
When you are ready to remove the sensor, make sure to pull out the
sensor pod while the transmitter is still attached.
1. Gently peel up the sensor pod adhesive patch from your skin, this will pull out
your sensor.
8.3
TRANSMITTER REMOVAL
Once the sensor pod is off your body, you will need to remove the transmitter (do
not dispose of the transmitter). To remove the transmitter you can use either
of the two methods below:
Method 1
The safety lock, once removed from the applicator (see Chapter 3, Section 3.4
Placing the Sensor), can be used as a tool to remove the transmitter.
1. Place the sensor pod on a table.
2. Hold the rounded edge of the safety lock.
3. Make sure the jagged edge of the safety lock is facing down (the direction
away from the removal arrow) as shown below:
Dexcom G4 User's Guide
|
95
Need help?
Do you have a question about the g4 and is the answer not in the manual?
Questions and answers
The mom set a high ceiling of 300. How do I get to an actual number and not a "high" reading
You can adjust the high glucose alert value on the Dexcom G4 by following these steps:
1. Press the button to select “High Alert.”
2. Press the UP or DOWN button to highlight “On/Off” and press the SELECT button to set this option.
3. Press the LEFT button to return to the previous screen.
4. Press the DOWN button to highlight “Level.”
5. Press the SELECT button and then use the UP or DOWN button to adjust the high glucose alert value.
6. The high glucose alert value can be set between 6.7 and 22.2 mmol/L in 0.1 mmol/L increments.
7. Press the SELECT button to confirm the new setting.
This adjustment allows you to set a specific threshold instead of receiving a general "high" reading.