11
Overall, the System reads, on average, 8.2% different (Mean Percent
Difference) than the YSI at glucose levels of 40-400 mg/dL (2.2-22.2 mmol/L)
and, 14.1% absolute different (Mean Absolute Difference) than the YSI values
at glucose levels of 40-400 mg/dL (2.2-22.2 mmol/L). The Median Percent
Difference shows that 50% of the time the System reads approximately 8.7%
more than the YSI blood glucose values and the Median Absolute Difference
shows that 50% of the time the System reads about 12.0% different than YSI
blood glucose values.
Clarke Error Grid Analyses
The Clarke Error Grid was used to determine if System results are "clinically
accurate." The Clarke Error Grid divides a correlation plot into 5 zones, which
are described in Table 3.
Table 3.
Clarke Error Grid Description
Zone
A
Clinically Accurate, would lead to correct treatment decisions
B
Would lead to benign decisions or no treatment
C
Would lead to overcorrection of normal glucose levels
D
Would lead to failure to be aware of and treatment of high or low glucose
levels
E
Would lead to inaccurate and "dangerous" treatment decisions
Results in zones A and B are considered "clinically acceptable," in that you
would most likely make accurate treatment decisions or decide not to treat with
the readings provided. Results in zones C, D, and E can lead to dangerous
treatment errors. The Clarke Error Grid zones are labeled on a correlation plot in
Figure 1. Each YSI result (horizontal axis) is matched up with its corresponding
System result (vertical axis). A point is recorded where the two readings
intersect. These results are based on 1477 paired data points recorded.
132
|
Dexcom G4 Continuous Glucose Monitoring System
Description
Need help?
Do you have a question about the g4 and is the answer not in the manual?
Questions and answers
The mom set a high ceiling of 300. How do I get to an actual number and not a "high" reading
You can adjust the high glucose alert value on the Dexcom G4 by following these steps:
1. Press the button to select “High Alert.”
2. Press the UP or DOWN button to highlight “On/Off” and press the SELECT button to set this option.
3. Press the LEFT button to return to the previous screen.
4. Press the DOWN button to highlight “Level.”
5. Press the SELECT button and then use the UP or DOWN button to adjust the high glucose alert value.
6. The high glucose alert value can be set between 6.7 and 22.2 mmol/L in 0.1 mmol/L increments.
7. Press the SELECT button to confirm the new setting.
This adjustment allows you to set a specific threshold instead of receiving a general "high" reading.
This answer is automatically generated