To use your Dexcom G4 Continuous Glucose Monitoring System
you will need a sensor, a transmitter, and a receiver. You will also
need a blood glucose meter and test strips for calibration. Once
inserted and calibrated, the sensor will continuously measure
and display your sensor glucose readings for up to 7 days. The
following sections will show you how to insert the sensor and
start a new continuous glucose monitoring session.
Please review the tutorial on the CD in your kit. The tutorial is
also available online at www.dexcom.com.
Sensors may fracture on rare occasions. If a sensor breaks and no
portion of it is visible above the skin, do not attempt to remove it.
Seek professional medical help if you have symptoms of infection or
inflammation—redness, swelling or pain—at the insertion site. If you
experience a broken sensor, please report this to your local distributor.
For patients undergoing an MRI with a retained wire broken off from a Dexcom
G4 Sensor, in-vitro MRI testing did not detect any safety hazards. There was no
significant migration or heating of the wire and imaging artifacts were limited to
the area around the wire.
3.1
BEFORE YOU START
r
Make sure you charge your receiver completely. See Chapter 1, Section 1.4,
Charging Your Receiver Battery.
r
Check that the date and time are correct on the receiver.
Make sure the correct transmitter ID has been entered into your receiver (see
r
Chapter 2, Section 2.2, The Settings Menu).
r
Check the Use by Date on the sensor. The Use by Date format is YYYY-
MM-DD. Sensors must be inserted on or before the end of the calendar day
printed on the sensor package label.
Dexcom G4 User's Guide
|
35
Need help?
Do you have a question about the g4 and is the answer not in the manual?
Questions and answers
The mom set a high ceiling of 300. How do I get to an actual number and not a "high" reading
You can adjust the high glucose alert value on the Dexcom G4 by following these steps:
1. Press the button to select “High Alert.”
2. Press the UP or DOWN button to highlight “On/Off” and press the SELECT button to set this option.
3. Press the LEFT button to return to the previous screen.
4. Press the DOWN button to highlight “Level.”
5. Press the SELECT button and then use the UP or DOWN button to adjust the high glucose alert value.
6. The high glucose alert value can be set between 6.7 and 22.2 mmol/L in 0.1 mmol/L increments.
7. Press the SELECT button to confirm the new setting.
This adjustment allows you to set a specific threshold instead of receiving a general "high" reading.