Dexcom g4 User Manual page 137

Continuous glucose monitoring system
Hide thumbs Also See for g4:
Table of Contents

Advertisement

Table 9.
Percentage of CGM System Reading within Paired SMBG Results
SMBG Glucose
Ranges mg/dL
(mmol/L)
40-400 (2.2-22.2)
*40-60 (2.2-3.3)
*61-80 (3.4-4.4)
81-180 (4.5-10.0)
181-300 (10.1-16.7)
301-350 (16.8-19.4)
351-400 (19.5-22.2
*For SMBG less than 80 mg/dL (4.4 mmol/L), the absolute difference is presented as the
difference between the System and SMBG, rather than the percent.
Table 10.
System and SMBG Glucose Accuracy
SMBG Glucose
Ranges mg/dL
(mmol/L)
System-SMBG
40-400 (2.2-22.2)
*40-60 (2.2-3.3)
*61-80 (3.4-4.4)
81-180 (4.5-10.0)
181-300 (10.1-16.7)
301-350 (16.8-19.4)
351-400 (19.5-22.2)
*For SMBG less than 80 mg/dL (4.4 mmol/L), the differences in mg/dL (mmol/L) are
included instead of percent differences (%).
Number
Percent
of Paired
within 20%
SMBG-CGM
of SMBG*
Readings
2824
83%
64
91%
249
81%
1726
83%
727
85%
42
93%
16
94%
Mean
Number
Percent
of Paired
Difference
(%)
2824
1.7
64
8.3 (0.5)
249
7.0 (0.4)
1726
2.4
727
-3.2
42
-7.6
16
-7.2
Percent
Percent
within
greater than
30% of
40% of SMBG*
SMBG*
95%
97%
95%
94%
96%
100%
100%
Mean
Median
Absolute
Percent
Percent
Difference
Difference
(%)
(%)
1.9
12.6
8.5 (0.5)
11.8 (0.7)
7.0 (0.4)
13.1 (0.7)
3.1
12.1
-2.6
11.0
-5.9
9.3
-5.4
7.8
Dexcom G4 User's Guide
2%
2%
3%
2%
1%
0%
0%
Median
Absolute
Percent
Difference
(%)
9.9
9.5 (0.5)
10.0 (0.6)
9.8
8.8
6.8
5.4
|
137

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the g4 and is the answer not in the manual?

Questions and answers

Kari Scott
March 27, 2025

The mom set a high ceiling of 300. How do I get to an actual number and not a "high" reading

1 comments:
Mr. Anderson
March 27, 2025

You can adjust the high glucose alert value on the Dexcom G4 by following these steps:

1. Press the button to select “High Alert.”
2. Press the UP or DOWN button to highlight “On/Off” and press the SELECT button to set this option.
3. Press the LEFT button to return to the previous screen.
4. Press the DOWN button to highlight “Level.”
5. Press the SELECT button and then use the UP or DOWN button to adjust the high glucose alert value.
6. The high glucose alert value can be set between 6.7 and 22.2 mmol/L in 0.1 mmol/L increments.
7. Press the SELECT button to confirm the new setting.

This adjustment allows you to set a specific threshold instead of receiving a general "high" reading.

This answer is automatically generated

Subscribe to Our Youtube Channel

Table of Contents