6. Do not use wipes that contain adhesives (e.g. Smith + Nephew IV Prep).
7. Place the device on a clean, dry cloth and air dry for 2-3 minutes.
Disinfection
Disinfection removes and destroys microorganisms and pathogens from the
surface of the device. Disinfect the receiver and transmitter periodically or
whenever you suspect that blood or body fluid has come in contact with the
surface of the device. If a second person, such as a healthcare provider, is
helping you operate the receiver or transmitter, the device should be disinfected
before the second person uses it. You will need gloves, clean, dry absorbent
wipes, 70% isopropyl alcohol wipes and several disinfectant wipes containing
a bleach solution strength of 6500 parts per million that are indicated to kill
viruses (such as Dispatch Hospital Cleaner Disinfectant Towels with Bleach or
equivalent).
Preparation:
1. Use precaution when handling products worn or handled by another person.
2. Wash hands thoroughly.
3. Wear personal protective equipment as appropriate (gloves, protective
goggles, gowns, etc.).
4. Close the receiver's USB port slide cover.
5. The receiver is not waterproof. Do not use a soaking wet cloth.
Disinfecting the receiver or transmitter:
1. Wear gloves.
2. Thoroughly pre-clean the surface of any visible contamination with one
disinfectant wipe (wipe the front, back and all four sides of the device).
3. Thoroughly wet the surface of the device with a second disinfectant wipe to
wipe the front, back and all four sides of the device.
4. The surface should remain wet for at least 1 minute at room temperature (21º
C) to ensure proper disinfection. Use additional disinfection wipes to make
sure the surface stays wet for the full minute.
5. Dry the unit with a clean, dry absorbent wipe.
6. Wipe the outside of the device with a 70% isopropyl alcohol wipe to remove
any disinfectant residue.
7. Dry the unit with a clean, dry absorbent wipe.
Dexcom G4 User's Guide
|
115
Need help?
Do you have a question about the g4 and is the answer not in the manual?
Questions and answers
The mom set a high ceiling of 300. How do I get to an actual number and not a "high" reading
You can adjust the high glucose alert value on the Dexcom G4 by following these steps:
1. Press the button to select “High Alert.”
2. Press the UP or DOWN button to highlight “On/Off” and press the SELECT button to set this option.
3. Press the LEFT button to return to the previous screen.
4. Press the DOWN button to highlight “Level.”
5. Press the SELECT button and then use the UP or DOWN button to adjust the high glucose alert value.
6. The high glucose alert value can be set between 6.7 and 22.2 mmol/L in 0.1 mmol/L increments.
7. Press the SELECT button to confirm the new setting.
This adjustment allows you to set a specific threshold instead of receiving a general "high" reading.
This answer is automatically generated