11
Receiver
Immunity Test
IEC 60601 Test
Level
Power Frequency
3 A/m
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
Note: UT is the a.c. main voltage prior to application of the test level.
Receiver
IEC 60601
Immunity Test
Test Level
Conducted RF
3 Vrms
150 kHz to
IEC 61000- 4-6
80 MHz
124
|
Dexcom G4 Continuous Glucose Monitoring System
Receiver
Compliance
Level
3 A/m
Compliance
Electromagnetic Environment
Level
Portable and mobile RF
communications equipment should
be used no closer to any part of the
receiver, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation Distance
d =1.2 P
1/2
d = 1.2 P
1/2
d = 2.3 P
1/2
3 V
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Electromagnetic
Environment Guidance
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment
Guidance
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the g4 and is the answer not in the manual?

Questions and answers

Kari Scott
March 27, 2025

The mom set a high ceiling of 300. How do I get to an actual number and not a "high" reading

1 comments:
Mr. Anderson
March 27, 2025

You can adjust the high glucose alert value on the Dexcom G4 by following these steps:

1. Press the button to select “High Alert.”
2. Press the UP or DOWN button to highlight “On/Off” and press the SELECT button to set this option.
3. Press the LEFT button to return to the previous screen.
4. Press the DOWN button to highlight “Level.”
5. Press the SELECT button and then use the UP or DOWN button to adjust the high glucose alert value.
6. The high glucose alert value can be set between 6.7 and 22.2 mmol/L in 0.1 mmol/L increments.
7. Press the SELECT button to confirm the new setting.

This adjustment allows you to set a specific threshold instead of receiving a general "high" reading.

This answer is automatically generated