Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The transmitter is intended for use in the electromagnetic environment specified
below. The customer or the user of the transmitter should ensure that it is used
in such an environment.
Immunity Test
IEC 60601 Test Level
Electromagnetic
± 6 kV Contact
Environment
± 8 kV Air
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
± 2 kV for power
Transient /burst
supply lines
IEC 61000-4-4
± 1 kV for input/output
lines
Surge
± 1 kV
differential mode
IEC 61000-4-5
± 2 kV
common mode
Voltage
< 5 % UT (>95 % dip
Dips, Short
in UT) for 0.5 cycle
Interruptions
40 % UT (60 % dip in
and Voltage
UT) for 5 cycles
Variations on
70 % UT (30 % dip in
Power Supply
UT) for 25 cycles
Input Lines
< 5% UT (>95% dip in
IEC 61000-4-11
UT) for 5 sec
Power
3 A/m
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
Transmitter
Electromagnetic
Compliance
Environment
Level
Guidance
± 6 kV Contact
Floors should be wood,
concrete or ceramic tile.
± 8 kV Air
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%
Not
applicable
Battery
operated
Not
applicable
Battery
Operated
Not
applicable
Battery
operated
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment
Dexcom G4 User's Guide
|
121

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the g4 and is the answer not in the manual?

Questions and answers

Kari Scott
March 27, 2025

The mom set a high ceiling of 300. How do I get to an actual number and not a "high" reading

1 comments:
Mr. Anderson
March 27, 2025

You can adjust the high glucose alert value on the Dexcom G4 by following these steps:

1. Press the button to select “High Alert.”
2. Press the UP or DOWN button to highlight “On/Off” and press the SELECT button to set this option.
3. Press the LEFT button to return to the previous screen.
4. Press the DOWN button to highlight “Level.”
5. Press the SELECT button and then use the UP or DOWN button to adjust the high glucose alert value.
6. The high glucose alert value can be set between 6.7 and 22.2 mmol/L in 0.1 mmol/L increments.
7. Press the SELECT button to confirm the new setting.

This adjustment allows you to set a specific threshold instead of receiving a general "high" reading.