19
Working with measurement tools and isotherms
Emissivity
Reflected apparent temperature
Object distance
Atmospheric temperature
Relative humidity

19.8.4 Procedure

Follow this procedure:
1. Push the Menu/Back button.
2. On the main menu, go to the Parameters button
3. On the Parameters menu, select the parameter that you want to change and push
the joystick.
4. Move the joystick up/down to change the value.
5. Push the joystick to confirm.
Note
Of the five parameters above, emissivity and reflected apparent temperature are the two most impor-
tant to set correctly in the camera.
See also:
For more information about parameters, and how to correctly set emissivity and reflected
apparent temperature, see 30 Thermographic measurement techniques, page 80.
#T559772; r.7877/7877; en-US
0.95
+20°C (+69°F)
1.0 m (3.3 ft.)
+20°C (+69°F)
50%
and push the joystick.
50