28 – Thermographic measurement techniques
3
Measure the radiation intensity (= apparent temperature) from the reflecting source using the
following settings:
Emissivity: 1.0
D
: 0
obj
You can measure the radiation intensity using one of the following two methods:
10589003;a2
Figure 28.3 1 = Reflection source
Note: Using a thermocouple to measure reflected apparent temperature is not recom-
mended for two important reasons:
A thermocouple does not measure radiation intensity
A thermocouple requires a very good thermal contact to the surface, usually by
gluing and covering the sensor by a thermal isolator.
28.2.1.1.2
Method 2: Reflector method
1
Crumble up a large piece of aluminum foil.
2
Uncrumble the aluminum foil and attach it to a piece of cardboard of the same size.
3
Put the piece of cardboard in front of the object you want to measure. Make sure that the side
with aluminum foil points to the camera.
4
Set the emissivity to 1.0.
152
Publ. No. T559597 Rev. a500 – ENGLISH (EN) – December 10, 2010