The following tables of error codes provide a quick reference of the descriptions and
probable causes of error codes.
Table 10. General Error Codes
Code
C12
C13
E30
E31
E32
E33
E38
E42
E44
E45
Table 11. Blood Pressure Error Codes
Code
C01
C02
C03
C04
C05
C06
C07
E10
E11
Description
Device outside operating temperature
range.
Low battery level.
Internal malfunction.
Internal malfunction.
Internal malfunction.
Internal malfunction.
Date and time not set.
Internal malfunction.
Internal malfunction.
Internal malfunction.
Description
Blood pressure reading cancelled by user. Retake blood pressure reading.
Unable to release cuff pressure.
Inflation too quick.
Air leak.
Unable to determine blood pressure.
Unable to determine blood pressure.
Internal NIBP error.
Cuff pressure limits exceeded.
Cuff pressure duration exceeded.
Corrective Action
Change ambient temperature.
Charge battery.
Contact Technical Service.
Contact Technical Service.
Contact Technical Service.
Contact Technical Service.
Set date and time (see "Date/Time Menu Options" on
page 25)
Contact Technical Service.
Contact Technical Service.
Contact Technical Service.
Corrective Action
Check tubing for kinks and connection integrity.
Check tubing and connections.
Check blood pressure cuff and tubing connections.
Check connections; restrict patient movement.
Check connections; restrict patient movement.
Device will power down.
Device will power down.
Device will power down.