Masimo Spo 2 Functional Check; Nellcor Spo 2 Functional Check - Welch Allyn Spot Vital Signs LXi Service Manual

Hide thumbs Also See for Spot Vital Signs LXi:
Table of Contents

Advertisement

32
Functional Overview
8. The device will exit CAL mode after four minutes.
9. Test all available thermometers for calibration verification at the current Calibration Set
Point Temperature. Proceed to the next Calibration Set Point Temperature, see
"Changing the Calibration Set Point
Changing the Calibration Set Point Temperature
To scroll from one set point to the next, press and hold the Temperature Selection button
until a beep is heard. The newly selected set point appears in the upper left corner of the
LCD display. The device's current temperature is displayed, starts to flash, and continues
flashing until the cavity reaches the equilibrium at the new set point.
Masimo SpO
Functional Check
2
Use the Masimo Tester to perform to functionally check the Masimo sensors.
1.
Orient the Masimo Tester such that the mini-D connector mates with the SpO
connector on Spot Vital Signs LXi.
2. Power on Spot Vital Signs LXi and confirm the SpO
81%± 3% and the pulse reading is 61 bpm ± 1 bpm.
3. Place the thumb and index finger on the gray buttons on either side of the Masimo
Tester connector, press the buttons firmly, and gently pull to remove the tester.
There is no way to change the functionality of the SpO
calibration, contact Technical Service.
Nellcor SpO
Functional Check
2
Use a Nellcor-approved SpO
1.
Orient the simulator such that the connector mates with the SpO
Vital Signs LXi.
2. Power on Spot Vital Signs LXi and confirm the SpO
81%± 3% and the pulse reading is 61 bpm ± 1 bpm.
3. Place the thumb and index finger on either side of the simulator connector, press the
buttons firmly, and gently pull to remove the simulator.
There is no way to change the functionality of the SpO
functioning properly, contact Technical Service.
Temperature".
simulator (SRC-MAX ) to check the SpO
2
Welch Allyn Spot Vital Signs LXi
reading in the Display Window is
2
module. If the SpO
is out of
2
2
functionality.
2
connector on Spot
2
reading in the Display Window is
2
module. If the SpO
is not
2
2
2