Table 7. External Devices Menu Options
Setting
Information System
Barcode Patient ID
Barcode Clinician ID
Mandatory Sign In
Clear on Send/Save:
Clear on Power Cycle:
Weight Scale
Wireless Module
Printer
Printer Paper
Table 8. Save Readings Menu Options
Setting
Save Mode
Auto Save Interval
Reading Full Action
Table 9. Change Local Defaults Options
Setting
Language
BP Units
Temperature Units
Height Units
Weight Units
Date Format
Time Format
Description
On or off. You must enable this option to send patient readings wired or wirelessly.
On or off. You must enable this option to send patient readings wirelessly.
On or off. You must enable this option to require mandatory sign in of a clinician.
Yes or no. You must enable this option to require the sign-in of the clinician.
Yes or no. You must enable this option to clear the Clinician ID after after sending or
saving the measurements.
Yes or No. You must enable this option to clear the Clinician ID after powering the Spot
Vital Signs LXi off.
On or off. Spot Vital Signs LXi connects to a scale and the weight shows on the display
window (see "Technical Overview" on page 61 for scale details).
None or DPAC. You must enable DPAC to send patient readings wirelessly. The
wireless radio is available as an accessory only with Welch Allyn Connex.
On or off.
Plain or labels. Only available if the Printer is enabled.
Description
Manual or automatic. If automatic, Spot Vital Signs LXi saves readings at a preselected
time interval. For either option, Spot Vital Signs LXi automatically saves the measured
parameters into memory before automatically powering off when not used for 30
minutes.
Changes the amount of time before automatically saving the current patient reading.
Only available if Automatic Save Mode is enabled.
Auto Overwrite, Ask Overwrite, Do not Overwrite. Spot Vital Signs LXi can save 50
patient readings in memory. Upon reaching reading 51, the device may automatically
overwrite reading 1, ask the user if he/she wants to overwrite reading 1, or disable the
ability to take another reading until at least one reading is erased.
Description
English, Dansk, Nederlands, Suomi, Français, Deutsch, Italiano, Norsk, Español,
Português, Svenska, or Chinese.
mmHg or kPa.
Fahrenheit (°F) or Celsius (°C).
Inches (in) or centimeters (cm).
Pounds (lb) or kilograms (kg).
Displays the date in one of the following styles:
• mm/dd/yyyy example: July 16, 2005 = 07/16/2005
• dd/mm/yyyy example: 16 July 2005 = 16/07/2005
Displays the time in one of the following styles:
• 12-hour example: 5:00 PM
• 24-hour example: 17:00
Functional Overview