Service Work Checklist - Welch Allyn Spot Vital Signs LXi Service Manual

Hide thumbs Also See for Spot Vital Signs LXi:
Table of Contents

Advertisement

46
Troubleshooting

Service Work Checklist

Spot Vital Signs
Model Number
Test
Unit SW Version
SPO2 SW Version
Thermometry SW Version
Unit Pressure @ 0 mmHg
Unit Pressure @ 50 mmHg
Unit Pressure @ 150 mmHg
Unit Pressure @ 250 mmHg
Test Voltage @ 5.5 V
Current Test – Sleep State
Current Test – Patient DAQ
Current Test – Valve Pump
Noise Level
Button Test
Interface Test
Leak Test
Dump Test
Inflation Test
Valve Control Test
Over Pressure Test
SPO2 Sensor - Masimo
SPO2 Sensor - Nellcor
Temperature Calibration Suretemp Plus
Temperature Calibration Braun Pro 4000
Temperature Cal Key Suretemp Plus
Radio Ping Test
Serial Number
BP Cycle Count
Pass or Fail
Technician
Test Data
Pass/Fail
Welch Allyn Spot Vital Signs LXi
Date
Test Specification
+/- 1 mmHg
+/- 1 mmHg
+/- 1.5 mmHg
+/- 2.0 mmHg
+/- 0.1 VDC
< 0.5 A
< 1.25 A
< 2.0 A
< 0.05 mmHg
Pass/Fail
Pass/Fail
<= 6mmhg in 15 seconds
< 10 Seconds
< 7 Seconds
Pass/Fail
296 to 329 mmHg
81% +/- 3% 61bpm +/- 1bpm
90% +/- 1% 60bpm +/- 1bpm
96.8 +/- 0.2F 105.8 +/- 0.2F
96.8 +/- 0.2F 105.8 +/- 0.2F
97.3 +/- 0.2 F
Pass