Maintenance and Service

SureTemp Plus Thermometer

Temperature Probe
Press down on the connector tab and slide the connector out of the port to remove the
temperature probe.
Regularly wipe the probe with a cloth dampened with warm water and a mild detergent
solution, a 70% isopropyl alcohol solution, or a 10% chlorine bleach solution.
Removable Probe Well
1.
Remove the temperature probe from Spot Vital Signs LXi (see "Temperature Probe"
on page 86).
2. Grasp the well under the probe opening and pull up gently to remove it from the
device.
3. Swab the inner and outer surface of the probe well with a cloth dampened with a mild
detergent solution, 70% isopropyl alcohol, or 10% chlorine bleach solution. Immerse
the probe well in mild detergent solution as necessary for cleaning.
4. Dry all surfaces thoroughly before re-assembling the device (see "SureTemp Plus" on
page 20) for reassembly instructions.

Braun ThermoScan PRO 4000 Thermometer

Use a soft cloth slightly moistened with alcohol to clean the thermometer display and
exterior. Do not use abrasive cleaners.
Damage to the probe window or the presence of dirt or cerumen on the probe window
can affect the accuracy of your temperature measurement. To clean the window, gently
wipe it with a cotton swab slightly moistened with alcohol and immediately wipe dry with
a clean cotton swab. Allow to dry at least five minutes before taking a temperature.
Every month, clean the Braun ThermoScan PRO 4000 charging contacts within the Spot
Vital Signs LXi housing with a swab slightly dampened with alcohol.
Caution DO NOT immerse or soak the probe in any type of fluid.
Caution DO NOT use steam, heat, or gas sterilization on the probe.
Caution DO NOT autoclave the probe.
Caution DO NOT use hard or sharp objects to clean the probe well. This could
damage the probe well and cause the device to not function properly.
Caution DO NOT use steam, heat, or gas sterilization on the probe well.
Caution DO NOT autoclave the probe well.