Temperature Specifications - Welch Allyn Spot Vital Signs LXi Service Manual

Hide thumbs Also See for Spot Vital Signs LXi:
Table of Contents

Advertisement

76
Specifications

Temperature Specifications

Temperature Range
SureTemp Plus
Braun ThermoScan PRO 4000
Calibration Accuracy
SureTemp Plus
Braun ThermoScan PRO 4000 for displayed
temperature ranges
Display Resolution
SureTemp Plus Predict Time
Oral
Adult Axillary
Pediatric Axillary
Rectal
Welch Allyn Spot Vital Signs LXi
80° to 110° F (26.7° to 43.3° C)
68° to 108° F (20° to 42.2° C)
±0.2° F (0.1° C) (Monitor Mode)
± 0.4° F (±0.2° C) 95.9° to 107.6° F (35.5° to 42° C)
±0.5° F (±0.25° C) (outside this temperature range)
Approx. 4 to 6 seconds
Approx. 12 to 15 seconds (age 18 years and older)
Approx. 10 to 13 seconds (age 17 years and younger)
Approx. 10 to 13 seconds
0.1° F or ° C