Service Manual
Mechanical
Dimensions
Weight
Mounting
Portability
Electrical
Power Requirements
Patents
6,827 ,488; 6,971,790; and other patents pending.
Battery
Sealed lead acid, with external charger.
The Spot Vital Signs LXi battery is 90 to 100% charged after 6 hours of charging. The
rechargeable batteries in the Braun ThermoScan PRO 4000 thermometer requires an
additional 1 hour to charge and the rechargeable batteries in the external printer require an
additional 4 hours to charge. The battery automatically charges when Spot Vital Signs LXi
is powered through the AC power transformer. An operator can use the device while the
battery is charging; however, the battery charges faster when the instrument is not in
operation.
Environmental
Operating Temperature
Storage/Transport Temperature
Relative Humidity
Operating Altitude
WARNING This device is not suitable for use in the presence of a flammable
anesthetic mixture with air or oxygen or nitrous oxide. An explosion may result.
May be hand-carried when held by the rear handle
Patient-rated isolation transformer is connected to AC mains:
100-240V, 50-60Hz, 0.4A or internal power: 6Vdc, 6Ah
Device with SureTemp Plus: -13° to 131°F (-25° to 55°C)
Device with Braun ThermoScan PRO 4000: -4° to 122°F (-20° to 50°C)
Specifications
Height: 10.63 " (27 cm)
Length/Braun: 8" (20.32 cm)
Length/SureTemp Plus: 7.5" (19 cm)
Depth: 5.25" (13.34 cm)
7.5 lbs (3.4 kg)
Self-supporting on rubber feet
Custom mobile stand
Custom wall mount
50° to 104° F (10° to 40° C)
15 to 95% (non-condensing)
-557 to 16,000 ft. (-170 to 4877 m)
79