Putting Into Operation - KERN ITS Operating Instruction

Platform scale
Hide thumbs Also See for ITS:
Table of Contents

Advertisement

ITS-BA-e-0720
1.3
1.3.1
Note
1.3.2
User manual ME-Nr.:22017077 01/07

Putting into operation

For startup, connect the terminal to an analog METTLER TOLEDO weighing platform
(see installation instructions METTLER TOLEDO Terminals IND4.. or call METTLER
TOLEDO Service).
Connecting the power supply
CAUTION!
Before connecting the scale to the mains, check whether the voltage value printed on
the rating plate corresponds with the local mains voltage.
▲ Never connect the device if the voltage value printed on the rating plate is different
to the local mains voltage.
➜ Plug the mains plug into the socket.
After connection, the device performs a self-test. When the zero display appears,
the device is ready to weigh.
➜ Calibrate the device in order to obtain the greatest possible precision, see
Section 4.3.2.
Terminals with AccuPac can work independently from the mains for approximately
30 hours in normal operation. A prerequisite for this is that the background lighting
is switched off and that no peripheral devices are connected.
The battery symbol indicates the present charging level of the battery. 1 segment cor-
responds to approx. 25 % capacity. When the symbol flashes the battery must be
charged (min. 4 hours). The charging period is extended if work is continued during
charging. The battery is protected against overcharging.
The battery's charging capacity can be reduced under continuous mains operation.
➜ To maintain the charging capacity, after a maximum of 4 weeks discharge the bat-
tery completely before recharging it.
Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight
must be checked at regular intervals within the scope of quality assurance. For this
purpose, the answerable user must define a suitable interval as well as the nature and
scope of this check. Information is available on KERN's home page (www.kern-
sohn.com) with regard to the monitoring of balance test substances and the test
weights required for this. Test weights and balances can be adjusted quickly and at
a reasonable price in KERN's accredited DKD calibration laboratory (return to national
normal).