ITS-BA-e-0720

6 Event and error messages

Error
Display Dark
Insufficient load
Overload
User manual ME-Nr.:22017077 01/07
Cause
Back lighting set too dark
No mains voltage
Unit switched off
Mains cable not plugged in
Brief fault
Load plate not on the scale
Weighing range not reached
Weighing range exceeded
Result not yet stable
Function not permissible
Zeroing not possible with over-
load or insufficient load
Average piece weight too low
No valid value from the reference
scale
No calibration
Average piece weight too low
Event and error messages
Remedy
➜ Set back lighting (b.LIGHt)
brighter
➜ Check mains
➜ Switch on unit
➜ Plug in mains plug
➜ Switch device off and back on
again
➜ Place load plate on the scale
➜ Set to zero
➜ Unload scale
➜ Reduce preload
➜ If necessary adjust vibration
adapter or weigh dynamically
➜ Unload scale and set to zero
➜ Unload scale
➜ Select and place larger number of
reference parts on the scale
➜ Check cable connection between
the units
➜ Check interface settings
➜ Unplug the mains plug then plug
it back in; switch unit off and then
back on in battery mode
➜ Calibrate scale
➜ Contact your dealer or local rep-
resentative
➜ Counting is not possible on this
scale with this average piece
weight
42