ITS-BA-e-0720
5.2
5.2.1
5.2.2
Status
1
2
3
4
STX
SWA SWB SWC MSD
Field 1
6 digits for the weight value that is sent without a decimal point and unit
Field 2
6 digits for the tare weight that is sent without a decimal point and unit
STX
ASCII characters 02 hex, characters for "start of text"
SWA, SWB, SWC
Status words A, B, C, see below
MSD
Most significant digit
LSD
Least significant digit
CR
Carriage Return, ASCII characters 0D hex
CHK
Checksum (2-part complement of the binary sum of the 7 lower bits of all previously sent
characters, incl. STX and CR)
User manual ME-Nr.:22017077 01/07
TOLEDO Continuous mode
TOLEDO Continuous commands
The scale supports the following input commands in TOLEDO Continuous mode:
Command
P <CR><LF>
T <CR><LF>
Z <CR><LF>
C <CR><LF>
Tx.xxx <CR><LF>
Output format in TOLEDO Continuous mode
Weight values are always sent in the following format in TOLEDO Continuous mode:
Field 1
5
6
7
8
–
–
–
Meaning
Print out the current result
Tare the scale
Zero the display
Clear the current value
Define tare
Field 2
9
10
11
12
–
LSD MSD –
Interface description
13
14
15
16
–
–
–
LSD CR
17
18
CHK
40