ITS-BA-e-0720
4.3.3
UNIt1
UNIt2
rESOLU
UNt.rOLL
Notes
4.3.4
A-tArE
CHAIn.tr
A.CL-tr
4.3.5
AZM
4.3.6
ON/OFF
User manual ME-Nr.:22017077 01/07
The scale calibrates with the calibration weight loaded. After calibration is completed,
-done- appeats briefly in the display.
Return to weighing mode:
Press the
-key, "END" will appear on the display
Press the
-key, order to save the alterations. "Save" will appear on the display.
Press
-key. to confirm. Following this the balance will return to weighing mode.
DISPLAY – weighing unit and display accuracy
Select weighing unit 1: g, kg, oz, lb, t
Select weighing unit 2: g, kg, oz, lb, t
Select readability (resolution), model-dependent
When UNT.rOLL is switched on, the weight value can be displayed in all available
units and as pieces with
•
On certified scales, the weighing units oz and lb are displayed with the symbol *.
•
On certified scales, resolutions that deviate from the scale definition are displayed
without a weighing unit and with the symbol *.
•
On dual-range/dual interval scales, resolutions marked with |<–> 1/2| are
divided up into 2 weighing ranges / intervals, e.g. 2 x 3000 d.
TARE – tare function
Switching on/off automatic taring
Switching on/off chain tare
Switching on/off automatic taring with automatic clearing of the tare weight when the
load is removed from scale
ZERO – automatic zero update
On certified scales, this menu item does not appear.
Switching on/off automatic zero update and selecting zeroing range.
Possible settings: OFF; 0.5 d; 1 d; 2 d; 5 d; 10 d
RESTART – automatic saving of zero point and tare value
When the Restart function is activated, the last zero point and tare value are saved.
After switching off / on or after a power interruption, the device continues to work with
the saved zero point and tare value.
.
Settings in the menu
29