Operating instructions
Platform scale
KERN IFB
Version 1.2
02/2010
GB
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
email: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
IFB-BA-e-1012

Advertisement

Table of Contents
loading

Summary of Contents for KERN IFB-BA-e-1012

  • Page 1 KERN & Sohn GmbH Ziegelei 1 Phone: +49-[0]7433- 9933-0 D-72336 Balingen Fax: +49-[0]7433-9933-149 email: info@kern-sohn.com Internet: www.kern-sohn.com Operating instructions Platform scale KERN IFB Version 1.2 02/2010 IFB-BA-e-1012...
  • Page 2: Table Of Contents

    KERN IFB Version 1.2 02/2010 Operating instructions platform scale Table of Contents Technical Data ....................4 Appliance overview ................... 7 Keyboard overview .......................8 2.1.1 Numeric input via navigation keys ..................9 Overview of displays.....................9 Basic Information (General) ................10 Proper use ........................10 Improper Use......................10...
  • Page 3 AUTO OFF Menu ......................... 25 Service, maintenance, disposal ..............27 Cleaning ........................27 Service, maintenance ....................27 Disposal........................27 Error messages ......................27 Data output RS 232C ................... 28 10.1 Technical Data......................28 10.2 Printer mode .......................29 10.3 Continuous data output ....................29 Instant help....................30 IFB-BA-e-1012...
  • Page 4: Technical Data

    Power pack secondary voltage 9V, 800mA Dimensions display unit 250 x 160 x 58 (B x D x H) mm Weighing surface mm 300 x 240 x 80 300 x 240 x 80 400 x 300 x 80 Weight kg (net) Interface RS 232 IFB-BA-e-1012...
  • Page 5 Power pack secondary voltage 9V, 800mA Dimensions display unit 250 x 160 x 58 (B x D x H) mm Weighing surface mm 400 x 300 x 80 400 x 300 x 80 500 x 400 x 80 Weight kg (net) Interface optional RS232 IFB-BA-e-1012...
  • Page 6 Power pack secondary voltage 9V, 800mA Dimensions display unit 250 x 160 x 58 (B x D x H) mm Weighing surface mm 500 x 400 x 80 650 x 500 x 80 650 x 500 x 80 Weight kg (net) Interface optional RS232 IFB-BA-e-1012...
  • Page 7: Appliance Overview

    2 Appliance overview Battery status display Keyboard Weight display Tolerance tag, see chpt. 7.6 Weighing unit Levelling screw Spirit level (underneath weighing platform) IFB-BA-e-1012...
  • Page 8: Keyboard Overview

    • Digit selection to the left Navigation key  • Calculate weighing data via interface • Delete • Switch-over gross weight net weight • Back to menu/weighing mode • Activate animal weighing function • Activate weighing with tolerance limits • Delete total added memory IFB-BA-e-1012...
  • Page 9: Numeric Input Via Navigation Keys

    Then select by using additional digits and change these by using Finish entry with 2.2 Overview of displays Display Significance Rechargeable battery very low STABLE Stability display ZERO Zero display GROSS Gross weight Net weight AUTO Automatic add-up enabled Weighing unit Adding Indicators for weighing with tolerance limits IFB-BA-e-1012...
  • Page 10: Basic Information (General)

    Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be...
  • Page 11: Basic Safety Precautions

    4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. 4.2 Personnel training The appliance may only be operated and maintained by trained personnel.
  • Page 12: Unpacking And Implantation

    6.2 Unpacking/implantation Take the balance carefully out of the package, remove the plastic cover and place it in the designated work area. Scope of delivery / serial accessories: • Balance, see chpt. 2 • Mains power supply • Operating instructions IFB-BA-e-1012...
  • Page 13: Mains Connection

    6.3 Mains connection Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN. 6.4 Adjustment...
  • Page 14 Carefully place adjusting weight in the centre of the weighing plate. Wait for stability display, then press After successful adjustment the appliance automatically returns to weighing mode. • An adjusting error or incorrect adjusting weight will be indicated by the error message, repeat adjustment procedure. IFB-BA-e-1012...
  • Page 15: Verification

    The legal regulation of the country where the balance is used must be observed! • Verification of the weighing system is invalid without the "seal". Notes on verified weighing systems Position for seals and verification switches 1. Self-destructing seal 2. Verification switch 3. Cover verification switch 4. Self-destructing seal IFB-BA-e-1012...
  • Page 16: Operation

    Overloading exceeding the stated maximum load (max) of the device, minus a possibly existing tare load, must be strictly avoided. The instrument may be damaged by overloading. Exceeding of maximum load is indicated by “----“ as well as a signal sound. Remove load from weighing system or reduce preload. IFB-BA-e-1012...
  • Page 17: Weighing With Tare

    The audio sound depends on the setting of the menu block “BEEP“. Options: • no Acoustic signal turned off • ok Acoustic signal sounds when load is within tolerance limits • ng Acoustic signal sounds when load is beyond tolerance limits IFB-BA-e-1012...
  • Page 18 ; the display used for entering the lower limit SET L appears. Press ; current setting appears. Use the navigation keys (See chpt. 2.1.1) for entering the lower limit, e. g. 1000 kg; the currently active digit will be flashing. Confirm input by Select using SET H IFB-BA-e-1012...
  • Page 19 Tare, if using a weighing container. Put on goods to be weighed, tolerance control is started • The tolerance control is not active when the weight is under 20d. • To delete limits enter a value of “00.000 kg“. IFB-BA-e-1012...
  • Page 20: Manual Totalizing

    This process can be repeated until the capacity of the weighing system is exhausted. Display of the saved weighing data: Press and the number of weighings followed by the total weight will be shown for 2 sec. To receive a printout, press during this display. IFB-BA-e-1012...
  • Page 21 Delete weighing data: Press at the same time. The data in the summation memory are deleted. Printout example: *************** 1.000kg *************** 0.500kg *************** 0.700kg *************** total 2.200kg First weighing Second weighing Third weighing Number of weighings / total IFB-BA-e-1012...
  • Page 22: Automatic Adding-Up

    Please note that the weighing system must be unloaded between the individual weighing procedures. This process can be repeated until the capacity of the weighing system is exhausted. Display and delete the weighing data, as well as printout examples see chpt. 7.7. IFB-BA-e-1012...
  • Page 23: Animal Weighing

    To cancel locking, keep pressed at the same time (2s) until “U LCK“ appears. 7.11 Display background illumination Keep pressed (3s) until “setbl“ appears. Press again and the current setting will be displayed. IFB-BA-e-1012...
  • Page 24: Automatic Switch-Off Function " Auto Off

    - current setting appears. to select desired setting. of on function disabled AUTO OFF of 5 Weighing system will switch off after 5 min Weighing system will switch off after 15 min of 15 Save entry by or cancel using IFB-BA-e-1012...
  • Page 25: Menu

    Use the navigation keys, see chpt. 2.1 to switch between Change settings the available settings. Acknowledge setting / Save entry by or cancel using exit the menu Return to weighing To exit menu, press repeatedly. mode IFB-BA-e-1012...
  • Page 26 DUAL Weighing scale type, capacity and readability (non- verified) or verification value (verified). Adjustment, see chpt. 6.5 Not documented P3 OTH Keyboard interlock enabled LOCK s. Kap. 7.9 / 7.10 Keyboard interlock disabled Animal weighing enabled Animal weighing disabled IFB-BA-e-1012...
  • Page 27: Service, Maintenance, Disposal

    • Do not apply aggressive detergents (solvents etc.). 9.2 Service, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply. 9.3 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
  • Page 28: Data Output Rs 232C

    • Use a suitable cable to connect the weighing balance to the interface of the printer. Faultless operation requires an adequate KERN interface cable. • Communication parameters (baud rate, bits and parity) of balance and printer must match.
  • Page 29: Printer Mode

    Space line <lf> Space line • Printout “summation memory“ ************** <lf> Space line TOTAL NO: Number weighing processes TOTAL wgt.: 0.447KG Total of all individual weighings ************** 10.3 Continuous data output con1: Weighing mode HEADER1: ST=STABLE, US=UNSTABLE HEADER2: NT=NET, GS=GROSS IFB-BA-e-1012...
  • Page 30: Instant Help

    • Great fluctuations in temperature. • Warm-up time was ignored. • Electromagnetic fields / static charging (choose different location/switch off interfering device if possible) Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer. IFB-BA-e-1012...