Calibration/Span Adjustment

Calibration is recommended after initial installation and each time the balance/scale
is moved.
Features
Calibration/adjustment can be
performed only when:
– there is no load on the balance/scale,
– the balance/scale is zeroed, and
– the internal signal is stable.
If these conditions are not met, an error
message is displayed.
Example: Calibrate/adjust span of the balance
Step
1. Switch on the balance/scale
2. Zero the balance/scale
3. Start calibration
The preset calibration weight
is displayed without the weight unit
(in this example, 5000 g)
4. To select a different calibration
weight value
5. Confirm calibration weight value and
adjustment
After the zero point is stored, the required
calibration weight flashes on the display.
6. Place the required calibration
weight on the balance/scale
The readout stops flashing if the weight
is applied within the defined time limit and
tolerance. If the weight value is accepted,
the display stops flashing and the stability
symbol
appears on the display.
7. Remove the calibration weight
8. Calibration/span adjustment has been completed
14
The weight required for calibration/
span adjustment is displayed.
Standard calibration weights for selected
models: to remove, see instructions on
page 6.
v
Press
to select a different weight
value.
To cancel the procedure: press and hold
the
key (> 2 sec.).
Enter
CF
(here: model ELT 6001)
/scale
Key (or instruction)
e
w
q
v repeatedly
Enter
CF
Display
start calibration/span