Only valid when the measuring regulations stipulated in the "Standard Operating Procedure" WKD-038 are maintained.
Model
Weighing
capacity
BP210D
210
BP300S
300 g
BP210d_a.xls
Model
Adjustment
weight
BP210D
internal
BP300S
internal
BP210d_a.xls
The following applies to these specifications: WKD-038 and Test Record WKD-038-1/09.96

Internal Span Adjustment

– Make sure that the menu code »1 9 3 - internal calibration/adjustment
accessible « is set (see page 27, "Balance Menu Code Settings)".
– Press the t key to tare the balance; then press q to activate the
calibration/adjustment function.
– The built-in, motorized calibration/adjustment weight is applied, then
removed automatically; the weight is stored; and the balance then returns
to the standard weighing mode.
– After completing the adjustment procedure, reset the » 9--1 « code in the
balance operating menu.
– Slide the menu access switch back to the left.
Replace the capped plug to close the opening.
Table of Adjustment Data
Readability
[d]
40
0,01
80
g
0,01
mg
0,1
0,1 mg
Span accuracy
Test
weight
200 g
300 g
Reproducility
Test
Permissible
weight
tolerances
(s)
20 g
0,02 mg
200 g
0,2 mg
Permissible
Tare
tolerances
weight
(±)
10/20
/30
0,4 mg
g
50/10
0/150
70/15
0,4 mg
g
0/230
SARTORIUS Basic
plus
Semimicro-Balances
Off-center load
Test
Permissible
weight
tolerances
(±)
200 g
0,4 mg E2
200 g
0,4 mg E2
Linearity
Test
Permissible
weight
tolerances
(±)
10
0,03
g
50
0,2
70
g
0,3 mg 1
TCS
mg 1
21