Application Programs

Toggling between Weight Units

With this application program you can toggle the display of a weight value back and forth
between two weight units (see table below).
Example: Toggle weight unit from pounds [lb] (application) to grams [g] (basic unit)
Step
1. Select application program
2. Select Toggling between Weight Units
3. Confirm unit
4. Select weight unit; in this example:
"5. Pounds" (see table below)
5. Confirm weight unit (pounds)
6. Place sample on balance/scale
7. Toggle weight unit
Menu code
1. user
1
)
2.grams
(fact. setting ELT models) Grams
4.carat
(fact. setting GD502)
5.Pound
2
)
6.ounce
2
)
7.tryo
2
)
8.tl.hon
2
)
9.tl.sin
2
)
10.tl.ta
2
)
11.grai
2
)
12.peny
2
)
15.tl.ch
2
)
22.pdoz
2
)
23.newt
2
)
1
) User-defined conversion can be loaded in balance with RS-232 or USB program option.
2
) Not on model GD502
8
Key (or instruction)
w > 2 sec
v
Enter
CF
v repeatedly
Enter
CF
v
Unit
Conversion factor
Grams
1.00000000000
1.00000000000
Carats
5.00000000000
Pounds
0.00220462260
Ounces
0.03527396200
Troy ounces
0.03215074700
Hong Kong taels
0.02671725000
Singapore taels
0.02645544638
Taiwanese taels
0.02666666000
Grains
15.4323583500
Pennyweights
0.64301493100
Chinese taels
0.02645547175
lb/oz
0.03527396200
Newtons
0.00980665000
Display
Display
o
g
o
lb
oz
ozt
tlk
tl
tl
GN
dwt
tl
lb:oz
N