Chapter 7
Component Removal/Installation
7.29 Power Switch
Overlay

Power switch overlay

position on front bezel
7-32
Follow these steps to remove the power switch overlay (Figure 7-30). Reverse
to install.
1. Open the UI (section 7.28).
2. From inside the front bezel, disconnect the power switch overlay cable
from the UI PCBA.
3. Carefully peel the power switch overlay from the front bezel.
4. Once the power switch overlay is removed, use isopropyl alcohol to
remove any remaining adhesive. Ensure that the bezel surface is clean
and dry.
CAUTION:
Do not flex the overlay.
NOTE:
When installing a new power switch overlay, peel the backing from the
overlay, feed the cable through the bezel cutout, align the overlay to the
front bezel, and press to adhere.
Power switch
overlay cable
Figure 7-30: Removing the Power Switch Overlay