7.14 Speakers

M3 x 6 screw (x2)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
M3 washer (x2)
1049766 Rev A
Follow these steps to remove the alarm speakers (Figure 7-14). Reverse to
install.
1. Remove front panel (section 7.10).
2. Disconnect speaker #2 from the power management (PM) PCBA.
(Speaker #1 is disconnected when the MC PCBA is disconnected.)
3. Remove the screw and washer that attaches each speaker to the base.
Speaker #2
(to PM PCBA)
V60 Ventilator Service Manual
Component Removal/Installation
Figure 7-14: Removing the Alarm Speakers
Chapter 7
Speaker #1
(to MC PCBA)
7-15