Philips Respironics V60 Service Manual page 77

Respironics v60 ventilator
Hide thumbs Also See for Respironics V60:
Table of Contents

Advertisement

Diagnostic
code
111C
DA PCBA ADC failed.
(CBIT)
111D
MC PCBA ADC failed.
(CBIT)
111E
CPU PCBA ADC failed.
(CBIT)
111F
Internal temperature high at CPU PCBA.
(CBIT)
1120
Internal temperature high at DA PCBA.
(CBIT)
1121
Internal temperature high at MC PCBA.
(CBIT)
1122
Blower temperature high.
(CBIT)
1123
Battery temperature high.
(CBIT)
1124
Battery failed.
(CBIT)
Audible and visual alarm indicators activate and the
ventilator continues to ventilate. Battery charger turns off.
1125
Cooling fan speed error.
(CBIT)
Ventilator overheating possible.
1126
Flash file system error.
1127
OVP circuit failed.
1049766 Rev A
Table 5-1: Diagnostic Codes and Troubleshooting
Description
V60 Ventilator Service Manual
Diagnostic Mode and Troubleshooting
Recommended repair
1.Replace DA PCBA.
1.Replace MC PCBA.
1.Replace CPU PCBA.
1.Replace CPU PCBA.
1.Verify that cooling fan is
operational.
2.Replace DA PCBA.
1.Verify that cooling fan is
operational.
2.Replace MC PCBA.
1.Verify that cooling fan is
operational.
2.Replace blower.
3.Replace MC PCBA.
1.Replace internal battery.
2.Replace PM PCBA.
1.Replace internal battery.
2.Replace PM PCBA.
1.Verify that cooling fan RPM is
4000 + 1000 RPM.
2.Replace fan.
3.Replace PM PCBA.
1.Repeat software download.
2.Replace CPU PCBA.
1.Replace MC PCBA.
Chapter 5
5-23