Required Test Equipment - Philips Respironics V60 Service Manual

Respironics v60 ventilator
Hide thumbs Also See for Respironics V60:
Table of Contents

Advertisement

8.1 Required Test
Equipment
1049766 Rev A
Table 8-3 summarizes the test equipment required, and Table 8-4 summarizes
the service accessories required for performance verification. Check the
calibration status of all test equipment before use.
Table 8-3: Required Test Equipment for Performance Verification
Description
Electrical safety analyzer
Pneumatic calibration analyzer(s)
capable of measuring low pressure
(cmH
O), flow rate (LPM), volume
2
(liters), and respiratory rate
Oxygen analyzer
Pressure analyzer capable of
measuring high pressure (PSI)
Digital multimeter (DMM) and
frequency counter
Test lung
Temperature/humidity monitor
NOTE:
An oxygen source capable of delivering 140 LPM (40-87 PSI) is
required for oxygen system tests.
V60 Ventilator Service Manual
Performance Verification
Recommended Manufacturer/Model
Dale LT 5440 or equivalent
TSI Certifier FA Plus (P/N 1040311),
or equivalent
TSI Certifier FA Plus or equivalent
TSI Certifier FA Plus or equivalent
Local supplier
IngMar QuickLung or equivalent
Fisher Scientific 11-661-14 or
equivalent
Chapter 8
8-5