7.16 Power Supply

EMI shroud
Power supply
Holes in EMI shroud fit over screw
heads on the side of the power supply
1049766 Rev A
Follow these steps to remove the power supply (Figure 7-16). Reverse to
install.
NOTE:
Do not remove any connections from the power supply terminal strip.
1. Remove MC PCBA (section 7.8) and PM PCBA (section 7.15).
2. Remove EMI shroud from power supply by slightly lifting the shroud
away from the power supply with a small screwdriver.
3. Remove the screws that attach the mounting plate to the base. (Power
supply includes attached mounting plate.)
4. Disconnect AC inlet cable and power harness.
NOTE:
Squeeze EMI shroud before reinstalling to ensure that it fits snugly over
the power supply.
Mounting plate
Figure 7-16: Removing the Power Supply
V60 Ventilator Service Manual
Component Removal/Installation
M3 x 6 screw (x4)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
From under mounting plate (not shown):
M3 x 6 screw (x4)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
Chapter 7
7-17