Specifications; Control Settings - Philips Respironics V60 Service Manual

Respironics v60 ventilator
Hide thumbs Also See for Respironics V60:
Table of Contents

Advertisement

Chapter 10.

10.1 Control settings

Table 10-1: Control setting ranges, resolutions, and accuracies
Parameter
Range
Modes
CPAP, S/T, PCV,
AVAPS (optional)
C-Flex
OFF, 1 to 3
CPAP
4 to 25 cmH
EPAP
4 to 25 cmH
IPAP
4 to 40 cmH
I-Time (Inspiratory
0.30 to 3.00 sec
Time)
Max P (AVAPS Maxi-
6 to 40 cmH
mum IPAP Pressure)
Min P (AVAPS Mini-
5 to 30 cmH
mum IPAP Pressure)
O
(Oxygen)
21 to 100%
2
Ramp Time
OFF, 5 to 45 min
Rate (Respiratory Rate)
4 to 60 BPM
Rise (Rise Time)
1 to 5
V
(AVAPS Target Tidal
200 to 2000 mL BTPS
T
Volume)
1049766 Rev A

Specifications

Table 10-1 lists ventilator control setting ranges, resolutions, and accuracies.
Table 10-2 lists the controls active in the different ventilation modes.
Resolution
Mode settings
N/A
Control settings
1
O
1 cmH
O
2
2
O
1 cmH
O
2
2
O
1 cmH
O
2
2
0.05 sec
O
1 cmH
O
2
2
O
1 cmH
O
2
2
1%
5 min
1 BPM
1
5 mL
V60 Ventilator Service Manual
Accuracy
Factory default
N/A
S/T
N/A
2
± (2 cmH
O + 4% of
4 cmH
2
target)
± (2 cmH
O + 4% of
4 cmH
2
target)
± (2 cmH
O + 4% of
12 cmH
2
target)
± 0.03 sec
1.00 sec
± (2 cmH
O + 4% of
25 cmH
2
target)
± (2 cmH
O + 4% of
10 cmH
2
target)
± 5%
21%
± 1 sec
OFF
± 1 BPM
4 BPM
N/A
3
± 15%
500 mL
O
2
O
2
O
2
O
2
O
2
10-1