7.32 Front Bezel,
Touch Screen
Nav-ring cable
Front bezel
M3 x 6 screw (x8)
(34 + 3.4 N cm/
3.0 + 0.3 in.-lbf)
Power switch
overlay cable
1049766 Rev A
Follow these steps to remove the front bezel and touch screen (Figure 7-33).
Reverse to install.
NOTE:
When installing a new touch screen, remove the protective film and avoid
touching the screen surface.
NOTE:
After installing a new UI assembly, perform the touch screen calibration
(section 5.3).
1. Open the UI (section 7.28).
2. Disconnect these cables from the UI PCBA:
Switch PCBA cable
Nav-ring cable
Power switch overlay cable
Touch screen cable
3. Remove the switch PCBA.
4. Remove the screw that attaches the cable clamp to the LCD tray.
5. Remove the screws and washers that attach the LCD tray to the front
bezel, then remove the LCD tray.
6. Remove the touch screen from the front bezel.
Switch PCBA to
UI PCBA cable
LCD tray
Cable clamp
M3 x 6 screw (x1)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
Touch screen
cable
Figure 7-33: Removing the Touch Screen
V60 Ventilator Service Manual
Component Removal/Installation
Chapter 7
Touch screen
7-35