1049766 Rev A
Figure 8-16: Diagnostics Menu - Pneumatics Tab Selected
7. Set ventilator pressure values and verify that they correspond to
pressure measurements:
NOTE:
Add the machine pressure offset reading (pneumatics screen) at 0 to the
analyzer readings for 1, 35, and 60 cmH
Set Pressure to:
0 cmH
O
2
1 cmH
O
2
35 cmH
O
2
60 cmH
O
2
8. To reset the pressure setting to 0, touch Flow then Accept.
9. The pressure accuracy test is complete. Remove low-pressure plug if
additional testing is not required.
V60 Ventilator Service Manual
Performance Verification
O.
2
Verify that analyzer, average Machine, and
Proximal displays read:
-1.5 to 1.5 cmH
0.5 to 1.5 cmH
2
32.5 to 38.5 cmH
54 to 66 cmH
2
Chapter 8
O
2
O
O
2
O
8-23