Philips Respironics V60 Service Manual page 81

Respironics v60 ventilator
Hide thumbs Also See for Respironics V60:
Table of Contents

Advertisement

Diagnostic
code
1216
Alarm message: Mask:x, Exh Port:y
Use Menu to change
Displays when ventilator is turned on. Displays selected
mask type and exhalation port.
1217
Alarm message: Power Has Been Restored
Power is restored following loss of power. The ventilator
restarts and continues ventilation in the mode set before
power was lost.
2000- 2003 Informational Diagnostic Codes (no user action required)
2000
System startup (diagnostic).
2001
System startup (ventilation).
2002
System shutdown.
2003
Restored default settings
5000
Passed high pressure leak test.
5001
Failed high pressure leak test.
Leak detected.
1049766 Rev A
Table 5-1: Diagnostic Codes and Troubleshooting
Description
5xxx: Leaks Tests Diagnostic Codes
V60 Ventilator Service Manual
Diagnostic Mode and Troubleshooting
Recommended repair
Select mask and port from Menu
tab. Message is removed when
user confirms selections, or after 5
minutes.
Check patient. Confirm that
ventilator and alarm settings are
appropriate.
Occurs when ventilator is power
ON in diagnostic mode: no action
required.
Occurs when ventilator is powered
ON in normal ventilation: no
action required.
Occurs when ventilator is powered
OFF: no action required.
Occurs when default settings are
restored: no action required.
No action required.
1.Verify that oxygen source is
adequate for this test.
2.Check for leaks at the oxygen
fitting, oxygen filter element
cap, oxygen solenoid valve, and
oxygen pressure transducer.
3.Check that the oxygen solenoid
valve is fully closed by verifying
that there is zero flow across the
oxygen flow sensor. If not,
replace the oxygen solenoid
valve.
4.Verify that the oxygen pressure
transducer is reading correctly.
If not, replace the DA PCBA.
Chapter 5
5-27