Chapter 4
Periodic Maintenance
4-6
19. Type
#CLRLOG
(all caps) in the dialog box, and press Enter to clear the diagnostic log
(Figure 4-5).
Figure 4-5: Clear Diagnostic Codes Command (HyperTerminal)
20. When the diagnostic log has been successfully cleared, the dialog box
displays:
?CLRLOGOK
21. Touch System Settings, and then touch Restore Default Settings.
22. Touch Date/Time and correct the date and time if necessary.
23. Turn the V60 ventilator off by pressing the ON/Shutdown button.
24. Upon successful completion of the performance verification, hours
reset, and restoring the default settings, enter the required
information on the PM and electrical safety labels.
V60 Ventilator Service Manual
1049766 Rev A