Reinitiate A Software Package - Philips Respironics V60 Service Manual

Respironics v60 ventilator
Hide thumbs Also See for Respironics V60:
Table of Contents

Advertisement

Appendix A
A.3 Reinitiate a
Software Package
A-8
9. Touch Service and verify the Options on the V60 ventilator are the
same as the options listed on the GatewayInterface screen.
10. Click Exit on the GatewayInterface screen.
11. Turn the V60 ventilator off by pressing the On/Shutdown button.
12. Option installation is complete.
Follow these steps to reinitiate a V60 ventilator software package using
Respi-Link.
NOTE:
Do not connect the null modem cable until instructed to do so.
1. Turn on the PC and verify the Gateway Interface is running by double-
clicking on Start Auto Respi-Link Gateway in the Respi-Link folder on
your desktop.
2. If the ventilator is not already in Diagnostic mode, press and hold the
nav-ring Enter button and turn on the ventilator by pressing the On/
Shutdown button on the user interface. Within 5 seconds, release and
press the nav-ring Enter button again to enter Diagnostic mode.
3. Connect a standard 9-pin male-female RS-232 null modem cable and
9-pin to 25-pin female-male adapter between the PC serial port and
the V60 ventilator.
4. Double click on the GatewayInterface icon to re-launch the install
screen (Figure A-11).
Figure A-11: GatewayInterface icon
V60 Ventilator Service Manual
1049766 Rev A