Returning Ventilator To Operation; Performance Verification Troubleshooting/Repair - Philips Respironics V60 Service Manual

Respironics v60 ventilator
Hide thumbs Also See for Respironics V60:
Table of Contents

Advertisement

8.6 Returning
Ventilator to
Operation
8.7 Performance
Verification
Troubleshooting/
Repair
1049766 Rev A
Follow these steps to return the ventilator to operation after a successful
performance verification:
1. Remove all test equipment, tools, and materials from ventilator.
2. Verify that the internal exhalation port orifice is not plugged.
3. If the ventilator is not already in diagnostic mode: power up the
ventilator: press and hold the nav-ring Enter button and the On/
Shutdown button on the user interface. Within 5 seconds of power up,
release and press the nav-ring Enter button to enter the Diagnostics
menu.
4. Touch System Settings, and then Restore Default Settings.
5. Touch Date/Time and verify that the date and time are correct.
6. Turn ventilator OFF.
Use the following troubleshooting procedures if the ventilator fails a
performance verification test. Make sure that the external measurement
devices such as the pneumatic calibration analyzer are calibrated and
functioning properly.
Perform the repair procedures in the order listed until the problem is resolved.
(See Chapter 5 for diagnostic mode procedures, and Chapter 7 for component
replacement procedures.)
CAUTION:
Troubleshooting and repair should be performed only by a qualified
service technician.
V60 Ventilator Service Manual
Chapter 8
Performance Verification
8-37