Annual Preventive Maintenance Instructions - Philips Respironics V60 Service Manual

Respironics v60 ventilator
Hide thumbs Also See for Respironics V60:
Table of Contents

Advertisement

Chapter 4
Periodic Maintenance
4.1 Annual
Preventive
Maintenance
Instructions
4-2
Follow these steps to perform the annual preventive maintenance (PM)
procedure on the V60 ventilator.
NOTE:
The annual PM procedure must be performed by a qualified service
technician in an appropriate setting, such as a biomedical repair facility.
Use a clean patient circuit and filters when performing any testing or
operational checkout.
1. Clean the exterior of the V60 ventilator using universal precautions
(including mask, gloves, eye protection).
2. Enter Diagnostic mode by pressing and holding the nav-ring Enter
button and turn on the ventilator by pressing the On/Shutdown button
on the user interface. Within 5 seconds, press the nav-ring Enter
button again to enter Diagnostic mode. The Diagnostics Menu appears
(Figure 4-1).
V60 Ventilator Service Manual
Figure 4-1: Diagnostics Menu
1049766 Rev A