Device Setup; Setup Overview; Placement; Connecting Ac Power - Philips Trilogy Evo Universal Instructions For Use Manual

Table of Contents

Advertisement

4. Device Setup

Setup Overview

To set up Trilogy Evo, follow the steps shown below. See the accompanying section for instructions.
Step
1.
Place the device.
2.
Connect AC power.
3.
Install filters.
4.
Connect a circuit.
5.
Add oxygen (optional).
6.
Start the device.

Placement

Place Trilogy Evo on a stable, flat, hard surface. Air must flow freely. Do not block the air vents with items such as bedding or curtains. Do not place
Trilogy Evo near any heating or cooling equipment or air supplies such as forced air vents, radiators, or air conditioners. Ensure the USB and
detachable battery panel doors remain closed when not in use.
If the device has been stored outside the normal operating temperature stated in "Technical Specifications," ensure the device reaches operating
temperature before connecting power.
See the "EMC Information" chapter for guidance on possible electromagnetic interference.

Connecting AC Power

Use the AC cord provided to connect AC power before pressing the power button. Verify Trilogy Evo Universal is using AC power, indicated by the
green LED light next to the power button.
To use other power sources, such as the detachable battery or an external battery, see the "Power Management" chapter.

Installing Filters

CBRN Filter
To install the CBRN filter:
1.
Ensure the CBRN filter adapter gasket is securely seated.
2.
Place the filter adapter onto the bayonet mount and twist clockwise to lock in place.
3.
Screw the CBRN filter onto the filter adapter.
Section
Placement
Connecting AC Power
Installing Filters
Connecting a Circuit
Adding Oxygen
Starting Trilogy Evo Universal
Installing Filters